Universal measuring device for magnetic fields

Teslameter M-Test MK4 or also called gaussmeter, rendered with measuring probe. Active display shows measured value, since the value is relatively high, an LED on the probe is lit. The colour corresponds to that of the measured pole.
  • Universal measuring device for measurements in the air gap or on hard-to-reach areas with < 1 mm thin tangential probe
  • Measures magnetic flux density of permanent magnets escaping from the surface
  • two switchable units: kA/m, mT
  • Magnetic fields measurable up to +/- 2 Tesla
  • automatic switching between direct and alternating magnetic fields

Universal

Measuring device used to measure magnetic fields

Static or dynamic magnetic fields are detected and measured precisely with the M-Test MK4. The < 1 mm thick tangential probe allows accurate, selective measurements at hard-to-reach places, in air gaps or on surfaces. The M-Test MK4 is especially suited for measuring permanent magnets or ferromagnetic small parts.

The immediately responding color LED directly indicates magnetism when the probe travels over the surface of the part. The “Peak Hold” function is controlled with the push button on the probe body; it makes it possible to easily find and define the places with the highest field strength for each polarity.

Technical data*

M-Test MK4
Measuring range, DC field–19,99 to +19,99 mT resp. –15,9 to +15,9 kA/m
–199,9 to +199,9 mT resp. –159 to +159 kA/m
–1999 to +1999 mT resp. –1590 to +1590 kA/m
(mT – kA/m selectable)
Measuring range, AC field14,14 mT resp. 11,25 kA/m
141,4 mT resp. 112,5 kA/m
1414 mT resp. 1125 kA/m True-RMS
frequency range 14…1000 Hz sine, triangle, rectangle
Accuracy in homogenous magnetic field+/– 1,5 % in low range; +/– 1 digit at 23 °C
+/– 0,6 % in mid range; +/– 1 digit at 23 °C
+/– 0,6 % in high range; +/– 1 digit at 23 °C
Zero-offset (after 1 min.)+/–  0,05 mT; +/–  0,04 kA/m
LED polarity display on the probered = north pole, green = south pole, orange = AC field
Polarity display on the gauge( ) = north pole, (–) = south pole
Threshold of polarity display (LED)+/– 0,5 mT; +/– 0,4 kA/m
Automatic OFF with battery power~4 min
Battery9 V, type 6 LF 22; AM6. Endurance~20 h
Length of probe cable1,2 m
Dimensions measuring deviceW84
(mm)H158
D30
Dimensions of probeW18
(mm)H22
D170
Dimensions sensorW8
(mm)H< 1
D70
Weight, including battery300g
Delivery includes
  • Indicating device
  • Operating manual
  • Battery
  • Soft storage case
Options**
  • Calibration certificate according to EN ISO 9001 and 10012
  • Reference magnet
  • Mini Zero Gauss Chamber
  • Hard case
  • Power adapter 9 V DC, power stabilized
  • Zero Gauss chamber 17“or 21“ for shielding from geomagnetic field

M-Test MK4

Measuring range, DC field –19,99 to +19,99 mT resp. –15,9 to +15,9 kA/m
–199,9 to +199,9 mT resp. –159 to +159 kA/m
–1999 to +1999 mT resp. –1590 to +1590 kA/m
(mT – kA/m selectable)
Measuring range, AC field 14,14 mT resp. 11,25 kA/m
141,4 mT resp. 112,5 kA/m
1414 mT resp. 1125 kA/m True-RMS
frequency range 14…1000 Hz sine, triangle, rectangle
Accuracy in homogenous magnetic field +/– 1,5 % in low range; +/– 1 digit at 23 °C
+/– 0,6 % in mid range; +/– 1 digit at 23 °C
+/– 0,6 % in high range; +/– 1 digit at 23 °C
Zero-offset (after 1 min.) +/– 0,05 mT; +/– 0,04 kA/m
LED polarity display on the probe red = north pole, green = south pole, orange = AC field
Polarity display on the gauge ( ) = north pole, (–) = south pole
Threshold of polarity display (LED) +/– 0,5 mT; +/– 0,4 kA/m
Automatic OFF with battery power ~4 min
Battery 9 V, type 6 LF 22; AM6. Endurance~20 h
Length of probe cable 1,2 m
Dimensions measuring device in mm (W x H x D) 84 x 158 x 30
Dimensions of probe in mm (W x H x D) 18 x 22 x 170
Dimensions sensor in mm (W x H x D) 8 x < 1 x 70
Weight, including battery 300g
Delivery includes
  • Indicating device
  • Operating manual
  • Battery
  • Soft storage case
Options
  • Calibration certificate according to EN ISO 9001 and 10012
  • Reference magnet
  • Mini Zero Gauss Chamber
  • Hard case
  • Power adapter 9 V DC, power stabilized
  • Zero Gauss chamber 17“or 21“ for shielding from geomagnetic field

* All informations are without guarantee